GAS LIQUID CHROMATOGRAPH (GC) Application For analysis of pesticides, Bench top, PC controlled to interface with data management system, Minimum two channel system with two columns, two injector and with two detectors(FPD & ECD) be supplied fitted.
Pressure (set point) 0 - 1 0 0 p s i with resolution of 0.001psi
Column Oven 01 number
Temp. range Up to 400 oC or better
Temp. programming steps / ramp 20 ramps
Injector System 02 numbers
Injection port system Split and split less
Option to install maximum injectors 2
Operating temp. (max.) 400C
Flow control unit 01 number
Maximum flow rate 0-1250 mL/ minute
Pressure range Upto100 psi
FPD (Plus) (Flame Photometric Detector) 01 number
Operating Temperature 400C
Detection Limit <45 fg P/s, <2.5 S/s, with methyl parathion
Dynamic Range >103 S, 104 P with methyl parathion
Selectivity 106g S/g C, 106 g P/g C
Data Acquisition Rate Upton 200 Hz
ECD (Electron Capture Detector) 01 number
Operating Temperature Up to 400C
Minimum Detectable Level <4.4 fg/mL
Linear Dynamic Range >5 x 104 with landane
Dynamic Range >5 x105with landane
Data Acquisition Rate Upto50 Hz
Radioactive Source <15 mCi 63Ni foil
Auto Sampler 01 number
Sample capacity 150 samples vials (1.5 ml capacity)
Accessories
Sample tray,
Spare 1000vials,
Spare ferrules,
Spare syringe
Data Management System Software Foreign branded latest PC core I7 with 19 inch LED monitor and laser printer as per SSC approved specifications with preloaded licensed MS Windows XP/Vista operating system loaded with compatible software for instrument control, acquisition and data processing and report customization with backup copies, Intergraded early maintenance feedback option, Intergraded method translation tool, integrated sleep & wake mode (to reduce gas and power consumption) software with the capacity to control GC through computer. Accessories
All standard accessories as per manufacturers standard including operation/service manual, certificate of country of origin/guarantee of good to be supplied by the firm as brand new (one set).
Hydrogen, Nitrogen, Helium Cylinders, Air Compressor, 10 KVA UPS, Stabilizer, Gas supply Pipes, Filters, spare glass inserts, tubing etc. (one each)
Moisture Trap for Air
Syringe; 1L, 5L, 10L and 50L (2 each) for manual injection, Auto injector syringe (10 L range) as per manufactures standard (05 each).
Capillary Column DB-5 ms,0.25mm x 30 x0.25m (04 Each)
Capillary Column DB-1 ms,0.25mm x 30 x0.25m (04 Each)
Septum10 pack (25pieces/pack) and Ceramic Flame Jet 10Nos.
Graphite Ferrules 10 packs ( 10 per pack)
Silica gel 02 packs ( 1kg/ pack)
Empty glass columns 05 Nos. ( 1.2mm )
Tool Kit 1No.
Digital Thermometer with probe 01No.
Glass insert for capillary and packed column (05each) xiv. Glass insert for capillary and packed column (05each)
Spare glass insert, tubing, operation manual Standard test samples for checking calibration at installation.
Country of Origin: USA, UK , Japan
Power Requirement: All systems operating at 220-230V AC 50/60 Hz.
Complete with: Branded computer compatible to operating software and Printer.
Warranty: Free;01 years for parts and service Extended 05 year for parts and service on payment
Note: All the accessories required for complete installation and commissioning /operation of system to be provided by the supplying firm along with onsite training of the staff.