LIQUIDCHROMA TO GRAPHY MASS SPECTROPHOTOMETER ( LC MS/MS)
Application / Type For Analysis of Pesticide residues, Bench top software control integrated high sensitivity with ESI and APCI probes
Solvent Delivery Module 01 number
Type Quaternary Gradient
Operating Pressure 5800psi
Flow rate Upton to 10 ml/minute
Flow rate precision 0.1%RSD
Degasser Online / Built in
Accessories As per Manufacturers standard Including installation /fitting kit And compatible on line Solvent Degasser and manual injector (20 loop)
Spares Piston/plunger seal, piston/plunger Seal replacement kit, piston /plunger for pump, prime drain valve inlet and outlet check valve, solvent bottles and Tray in line filter and injector seal.
Sample injection volume 0.1 to 90 l (standard) or better
Injection volume precision 0.25 % RSD
Mass Spectrometric Detector 01 number
Type True Tandem mass spectroscopy system must have thermal focusing beam collimation
Mass range Upton 2250 m/z
MRM Sensitivity: Signal-to- Noise ratio (S/N) ESI positive 1 pg of reserpine injected on column, 4quantifying on the transition m/z 609 to 195 S/N > 6000:1 Noise 1 RMS. ESI negative 1 pg of chloramphenicol injected on column, quantifying on the transition m/z 321 to 152 S/N > 2,000:1 Noise 1 RMS