GAS LIQUID CHROMATOGRAPH ( GC) Gas Chromatograph with following detectors.
FID (`` Flame Ionization Detector ).
NPD (Nitrogen, Phosphorus Detector
Application For analysis of pesticides Residues, Bench top, PC controlled to interface with data management system, Minimum two channel system with two columns, two injectors ( packed and Capillary) system with detector be supplied fitted.
Pressure (set point) 0 - 1 0 0 p s i with resolution of 0.001psi
Column Oven 01
Temp. range Up to 450C
Temp. programming steps / ramp 20 ramps
Injector System 02
Injection port system Split and split less
Option to install maximum injectors 2
Operating temp.(max.) 400C
Flow control unit
Maximum flow rate 0-1250 mL/ minute
Pressure range Upto100 psi
FID (Flame ionization detector) 01
Operating Temperature Up to 450C
Detection limit <1.4pg C/sec
Dynamic range / linearity >107
Data Acquisition Rate Upton 500 Hz
Accessories
Including sample tray, spare 1000 vials, spare ferrules, spare syringe
Data Management System Software Foreign branded latest PC core I7 with 19 inch LED monitor and laser printer as per SSC approved specifications with preloaded licensed MS Windows XP/Vista operating system loaded with compatible software for instrument control, acquisition and data processing and report customization with backup copies, intergraded early maintenance feedback option, Intergraded method translation tool, integrated sleep & wake mode (to reduce gas and power consumption) software with the capacity to control GC through computer. Accessories
All standard accessories as per manufacturers standard including operation/service manual, certificate of country of origin/guarantee of good to be supplied by the firm as brand new (one set).
Hydrogen Generator , generating hydrogen 99.9999% pure H2, 250 cc per hour, with spare cartridges ,
Hydrogen, Nitrogen, Helium Cylinders, Air Compressor, 7KVA UPS, Stabilizer, Gas supply Pipes, Filters etc. (one each)
Moisture Trap for Air
Syringe; 1L, 5L, 10L and 50L (2 each) for manual injection,
Auto injector syringe (10 L range) as per manufactures standard (05 each).
Capillary Column DB-5 ms,0.25mm x 30 x0.25m (04 Each)
Capillary Column DB-1 ms,0.25mm x 30 x0.25m (04 Each)
Septum10 pack (25pieces/pack) and Ceramic Flame Jet 10Nos.
Graphite Ferrules 10 packs (10 per pack)
Silica gel 02 packs ( 1kg/pack)
Empty glass columns 05 Nos. (1.2mm)
Tool Kit 1No.
Digital Thermometer with probe 01No.
Glass insert for capillary and packed column (05each)
Spare glass insert , tubing,operation manual
Standard test samples for checking calibration at installation.
Computer Branded computer compatible with software and one printer (one number).
Power Requirement All system operating at 220-230 V AC 50/60 Hz
Country of Origin USA, UK , Japan
Warranty Free; 01 years for parts and service.
Extended; 05 year for parts and service on payment.
Note All the accessories required for complete installation and commissioning /operation of system to be provided by the supplying firm along with onsite training of the staff.